PXI-2597 Specifications
- Updated2023-02-18
- 6 minute(s) read
PXI-2597 Specifications
This document lists specifications for the PXI-2597. All specifications are subject to change without notice.
About These Specifications
Specifications characterize the warranted performance of the instrument under the stated operating conditions. Data in this document are Specifications unless otherwise noted.
Typical Specifications are specifications met by the majority of the instrument under the stated operating conditions and are tested at 23 °C ambient temperature. Typical specifications are not warranted.
All voltages are specified in DC, ACpk, or a combination unless otherwise specified.
Input
RF Performance
Characteristic impedance (Z0) | 50 Ω, nominal | ||||||||||||
| |||||||||||||
| |||||||||||||
| |||||||||||||
|
Dynamic
Relay operate/release time[4] | 15 ms |
Recommended cycle speed | 5 channels/s |
Expected mechanical relay life | 2 × 106 cycles |
Insertion loss repeatability | <0.03 dB, typical |
Trigger
| |||||||
|
Physical
Relay manufacturer/PN | Radiall/R574 series | ||||||||
Relay type | Electromechanical, non-latching | ||||||||
Contact material | Beryllium copper, gold-plated | ||||||||
I/O connector | 7 SMA jacks | ||||||||
SMA torque | 0.8 N · m to 1.1 N · m(7 in. · lbs to 10 in. · lbs) | ||||||||
| |||||||||
Dimensions (L × W × H) | 3U, three slot, PXI/cPCI module, 21.6 cm × 6.1 cm × 13.0 cm(8.5 in. × 2.4 in. × 5.1 in.) | ||||||||
Weight | 490 g (17.3 oz) |
Environment
Maximum altitude | 2,000 m (at 25 °C ambient temperature) |
Pollution Degree | 2 |
Indoor use only.
Operating Environment
Ambient temperature range | 0 °C to 55 °C (Tested in accordance with IEC 60068-2-1 and IEC 60068-2-2.) |
Relative humidity range | 10% to 90%, noncondensing (Tested in accordance with IEC 60068-2-56.) |
Storage Environment
Ambient temperature range | -40 °C to 70 °C (Tested in accordance with IEC 60068-2-1 and IEC 60068-2-2.) |
Relative humidity range | 5% to 95%, noncondensing (Tested in accordance with IEC 60068-2-56.) |
Shock and Vibration
Operational shock | 30 g peak, half-sine, 11 ms pulse (Tested in accordance with IEC 60068-2-27. Test profile developed in accordance with MIL-PRF-28800F.) | ||||||
|
Compliance and Certifications
Safety Compliance Standards
This product is designed to meet the requirements of the following electrical equipment safety standards for measurement, control, and laboratory use:
- IEC 61010-1, EN 61010-1
- UL 61010-1, CSA C22.2 No. 61010-1
Electromagnetic Compatibility
- EN 61326-1 (IEC 61326-1): Class A emissions; Basic immunity
- EN 55011 (CISPR 11): Group 1, Class A emissions
- EN 55022 (CISPR 22): Class A emissions
- EN 55024 (CISPR 24): Immunity
- AS/NZS CISPR 11: Group 1, Class A emissions
- AS/NZS CISPR 22: Class A emissions
- FCC 47 CFR Part 15B: Class A emissions
- ICES-001: Class A emissions
Product Certifications and Declarations
Refer to the product Declaration of Conformity (DoC) for additional regulatory compliance information. To obtain product certifications and the DoC for NI products, visit ni.com/product-certifications, search by model number, and click the appropriate link.
Environmental Management
NI is committed to designing and manufacturing products in an environmentally responsible manner. NI recognizes that eliminating certain hazardous substances from our products is beneficial to the environment and to NI customers.
For additional environmental information, refer to the Engineering a Healthy Planet web page at ni.com/environment. This page contains the environmental regulations and directives with which NI complies, as well as other environmental information not included in this document.
EU and UK Customers
电子信息产品污染控制管理办法(中国RoHS)
2 NI recommends against switching active RF signals. As a relay actuates, the channel is momentarily unterminated. Some RF sources can be damaged by reflections if their outputs are not properly terminated. Refer to your RF source documentation for more information.
3 National Instruments recommends against switching RF signals below -35 dBm with this device.
4 Certain applications may require additional time for proper settling. Refer to the NI Switches Help for more information about including additional settling time.
5 The PXI-2597 can recognize trigger pulse widths less than 150 ns if you disable digital filtering. Refer to the NI Switches Help for information about disabling digital filtering.