NI-TClk Manual Calibration Using NI-RFSA
- Updated2024-08-16
- 1 minute(s) read
NI-TClk Manual Calibration Using NI-RFSA
Complete the following steps to improve the synchronization accuracy by lowering skew and minimizing jitter of NI-TClk synchronized devices using manual calibration.
- Connect the external signal to the VSTs. For best performance, minimize unnecessary skew by using matched length cables.
- Configure the VST for acquisition using NI-TClk. If using LabVIEW, use the RFSA Synchronization (TClk, Shared LO and Reference Clock) (5840).vi example.
- Start the acquisition and measure the skew between devices, noting the amount of jitter.
- Read the value from the Absolute Delay property. Stop the acquisition, modify the value, and write it back to apply skew correction.
- Repeat steps 3 and 4 until you achieve the appropriate synchronization results.
- Read the value from the Absolute Delay property and store the value for future synchronization sessions.