2-Wire 16×16 Matrix

Software name: 2532/2-Wire 16x16 Matrix (NISWITCH_TOPOLOGY_2532_2_WIRE_16X16_MATRIX)

Hardware Diagram

The NI TB-2645B terminal block creates a 2-Wire 16×16 matrix topology with the module.


1378

Making a Connection

For bank 0, both the scanning command, r6->c1;, and the immediate operation, niSwitch Connect Channels VI or the niSwitch_Connect function with parameters r6 and c1, result in the following connection:

  • signal connected to R6+ is routed to C1+
  • signal connected to R6- is routed to C1-
Note While you can, for example, connect B0R1 to B0C0, you cannot connect B0R1 directly to B1C1 in this topology.

Terminal Block Connections

The NI TB-2645B terminal block connects banks of the module to create the 2-Wire 16×16 matrix topology.


1378

Column Pin Connections

The following table refers to the J3 connector on the NI TB-2645B.

Pin Number Column Pin Number Column Pin Number Column Pin Number Column
Pin 1 C0+ Pin 9 C4+ Pin 17 C8+ Pin 25 C12+
Pin 2 C0– Pin 10 C4– Pin 18 C8– Pin 26 C12–
Pin 3 C1+ Pin 11 C5+ Pin 19 C9+ Pin 27 C13+
Pin 4 C1– Pin 12 C5– Pin 20 C9– Pin 28 C13–
Pin 5 C2+ Pin 13 C6+ Pin 21 C10+ Pin 29 C14+
Pin 6 C2– Pin 14 C6– Pin 22 C10– Pin 30 C14–
Pin 7 C3+ Pin 15 C7+ Pin 23 C11+ Pin 31 C15+
Pin 8 C3– Pin 16 C7– Pin 24 C11– Pin 32 C15–
Note The column connection board J2 connector on the NI TB-2645 is not used in this configuration.

Row Pin Connections

The NI TB-2645B provides two ribbon cable headers for row connection. Use one cable header to connect to your application. Use the other cable header for column expansion.

The following table lists the pin assignments for row connection.

The following table refers to the J4 and J5 connectors on the NI TB-2645B.

Pin Number Row
1 R0+
2 R0–
3 R1+
4 R1–
5 R2+
6 R2–
7 R3+
8 R3–
9 R4+
10 R4–
11 R5+
12 R5–
13 R6+
14 R6–
15 R7+
16 R7–

The following table refers to the J6 and J7 connectors on the NI TB-2645B.

Pin Number Row
1 R8+
2 R8–
3 R9+
4 R9–
5 R10+
6 R10–
7 R11+
8 R11–
9 R12+
10 R12–
11 R13+
12 R13–
13 R14+
14 R14–
15 R15+
16 R15–

Row Protection Bypass Header

Each row signal is isolated from the reed relays through a 100 Ω resistor. To bypass this resistor, install a jumper on the appropriate pins of the header indicated in the table below. The following table lists the appropriate jumper position for bypassing the resistor on each row signal.

Bypassed Row NI TB-2645B
Header Pins Connected
R0+ J14 1-2
R0- J14 3-4
R1+ J14 5-6
R1- J14 7-8
R2+ J14 9-10
R2- J14 11-12
R3+ J14 13-14
R3- J14 15-16
R4+ J15 1-2
R4- J15 3-4
R5+ J15 5-6
R5- J15 7-8
R6+ J15 9-10
R6- J15 11-12
R7+ J15 13-14
R7- J15 15-16
R8+ J16 1-2
R8- J16 3-4
R9+ J16 5-6
R9- J16 7-8
R10+ J16 9-10
R10- J16 11-12
R11+ J16 13-14
R11- J16 15-16
R12+ J17 1-2
R12- J17 3-4
R13+ J17 5-6
R13- J17 7-8
R14+ J17 9-10
R14- J17 11-12
R15+ J17 13-14
R15- J17 15-16