Troubleshooting
- Updated2023-02-17
- 1 minute(s) read
Troubleshooting
ITS-16200 Does Not Power On
- Check the power source connection and voltage supply.
- Verify that the main power switch is turned to the On position.
- Confirm that the EPO button is not engaged.
- Verify that the main breaker is in the On position on the PEP.
- Verify that the temperature controller on the EPO panel has not detected a temperature higher than the programmed shutoff value.
- Check the incoming power lights on the rear side of the PDU.
- Verify that the PDU main breaker is on and that the control breaker is depressed.
- Check that the cables in and out of the PDU are connected.
- Verify the shutoff thermostats have been reset if the ITS-16200 shut down due to thermal overload.
PXI Chassis Does Not Power On
- Confirm that the EPO button is not engaged.
- Remove the brush panel and 3U and 1U filler panels to check that the PXI controller is properly seated inside the PXI chassis.
- Check the PXI chassis LED indicators.
- Check if the chassis power supply cable is connected.
SLSC Chassis Does Not Power On
- Confirm that the EPO button is not engaged.
- Open the rear panel door and check the SLSC chassis LED indicators.
- Verify the PS-17 is on.
- Use a DMM to confirm the PS-17 has an output of 24 V.
- Check if the chassis power supply cable is connected.