Troubleshooting

ITS-16200 Does Not Power On

  • Check the power source connection and voltage supply.
  • Verify that the main power switch is turned to the On position.
  • Confirm that the EPO button is not engaged.
  • Verify that the main breaker is in the On position on the PEP.
  • Verify that the temperature controller on the EPO panel has not detected a temperature higher than the programmed shutoff value.
  • Check the incoming power lights on the rear side of the PDU.
  • Verify that the PDU main breaker is on and that the control breaker is depressed.
  • Check that the cables in and out of the PDU are connected.
  • Verify the shutoff thermostats have been reset if the ITS-16200 shut down due to thermal overload.

PXI Chassis Does Not Power On

  • Confirm that the EPO button is not engaged.
  • Remove the brush panel and 3U and 1U filler panels to check that the PXI controller is properly seated inside the PXI chassis.
  • Check the PXI chassis LED indicators.
  • Check if the chassis power supply cable is connected.

SLSC Chassis Does Not Power On

  • Confirm that the EPO button is not engaged.
  • Open the rear panel door and check the SLSC chassis LED indicators.
  • Verify the PS-17 is on.
  • Use a DMM to confirm the PS-17 has an output of 24 V.
  • Check if the chassis power supply cable is connected.