Repair
- Updated2023-02-17
- 4 minute(s) read
Repair
Note Ensure that Data Record
System repair procedures are performed only by
qualified personnel.
Type | NI Part Number | Function and Model |
---|---|---|
Chassis | 787976-01 | PXIe-1092DC, 10-slot (7 hybrid slots) 3U PXI Express chassis |
787975-01 | PXIe-1095DC, 18-slot (5 hybrid slots) 3U PXI Express chassis | |
Controller | 785831-01 | PXIe-8861, 2.8 GHz quad-core Intel Xeon PXI controller |
787567-01 | PXIe-8861 non-TPM (for China) | |
786777-01 | 8 GB DDR4-2133 SO-DIMM RAM, non-ECC, for PXIe-8861 | |
783513-05 | PXIe-8880, 2.3 GHz eight-core Intel Xeon PXI controller | |
783513-0518 | PXIe-8880 non-TPM (for China) | |
783813-01 | 8 GB Upgrade/Replacement RAM for PXIe-8880 | |
787807-01 | PXIe-8881, 3.9 GHz eight-core Intel Xeon PXI controller | |
787807-0118 | PXIe-8881 non-TPM (for China) | |
787659-01 | 16 GB DDR4 2666 SO-DIMM RAM, ECC for PXIe-8881 | |
Power Supply | 861354-01 | Data Record System AD 500W DC power supply |
Camera Interface | 787453-01 | PXIe-1486, FPD-Link III 8-input channel |
787454-01 | PXIe-1486, FPD-Link III 4-channel tap | |
787456-01 | PXIe-1487, GMSL 8-input channel | |
787457-01 | PXIe-1487, GMSL 4-channel tap | |
783625-01 | PXIe-7976, 2 GB DRAM PXI FPGA instrument for FlexRIO, K410T FPGA, 3.2 GB/s | |
787313-01 | PXIe-8245, 1 Gigabit Ethernet, 4 port | |
787314-01 | PXIe-8246, power over Ethernet, 1 Gigabit Ethernet, 4 port | |
Automotive Bus Interface | 783639-01 | PXIe-6592R, 10 Gbps high-speed serial, 4 channel |
784976-01 | PXIe-8238, 10 Gigabit Ethernet, 2 port | |
784977-01 | PXIe-8240, 40 Gigabit Ethernet, 2 port | |
787313-01 | PXIe-8245, 1 Gigabit Ethernet, 4 port | |
787314-01 | PXIe-8246, power over Ethernet, 1 Gigabit Ethernet, 4 port | |
784122-01 | PXIe-8510, X-NET interface, 2 port | |
784121-01 | PXIe-8510, X-NET interface, 6 port | |
784122-01 | PXI-8517, FlexRay interface, 2 port | |
786771-01 | PXIe-8521, automotive Ethernet interface instrument, 100Mbit, 4 port | |
787312-01 | PXIe-8522, automotive Ethernet interface instrument, Gigabit, 4 port | |
787311-01 | PXIe-8523, automotive Ethernet interface instrument, 100/1000BASE-T1, 4 port | |
787992-01 | PXIe-8623, 100/1000BASE-T automotive Ethernet interface instrument, 4 port | |
787104-01 | NI-9962, 3-pin push-in spring terminal (4 count) for automotive Ethernet | |
Timing and Synchronization | 782110-01 | PXI-6683H, TCXO (GPS, IRIG-B, IEEE 1588) PXI synchronization |
196304-30 | GPS antenna | |
779674-01 | SMA-SMB adapter | |
188859-01 | SMB-SMB coaxial cable | |
Data Storage | 787850-01 | PXIe-8267, 15 TB SSD, data storage (5 GB/s), 1 slot |
785157-01 | PXIe-8394, x8 Gen 3 MXI-Express daisy chain interface | |
Cables | 785550-01 | MXI-Express cable, Gen 3 x8, copper, 1 m |
785550-03 | MXI-Express cable, Gen 3 x8, copper, 3 m | |
785549-03 | MXI-Express cable, Gen 3 x4, copper, 3 m | |
783701-02 | TRC-8543, NI-XNET CAN transceiver cable | |
783702-03 | TRC-8546, NI-XNET LIN transceiver cable | |
Miscellaneous | 199198-01 | Assembly, slot blocker, 4HP, 3U, PXI |
185365D-01 | Assembly, filler panel | |
766489-01 Littelfuse PN: 153.5631.5801 |
DC power supply fuse | |
773765-01 Phoenix Contact PN: 1967456 |
External power connector |
Note This table is not exhaustive. Consult your
system integrator for a full list of parts specific to your system. Submit a request to NI for
any additional replacement parts your system requires.
Replacing the DC Power Supply Fuse
Equipment
- #1 Phillips screwdriver
- #2 Phillips screwdriver
- Verify the input power connector is disconnected from the Data Record System.
-
Unscrew the four screws securing the fuse cover to the DC power supply using
the #1 Phillips screwdriver.
- Fuse Cover Screws
- Fuse Cover
- Remove the fuse cover.
- Unscrew the two fuse screws using the #2 Phillips screwdriver.
-
Remove the fuse screws, lock washers, and fuse from the DC power supply.
- Fuse Screws
- Lock Washers
- DC Power Supply Fuse
- Place the new fuse (NI part number 766489-01) in the DC power supply.
- Secure the fuse with the two lock washers and two fuse screws using the #2 Phillips screwdriver.
- Torque the fuse screws to 0.76 N · m (6.7 lb · in.).
- Replace the fuse cover.
- Torque the fuse cover to the DC power supply using the #1 Phillips screwdriver to 0.56 N · m (5.0 lb · in.).
Replacing PXI Instruments
This topic contains general installation instructions for installing PXI instruments in the chassis. For specific instructions for your instrument, refer to the instrument documentation.
Equipment
- #1 Phillips screwdriver
- Connect the DC power source to the chassis.
- Ensure that the chassis is powered off.
- Disconnect all cables from the PXI instrument.
- Unscrew the captive screws on the PXI instrument using the #1 Phillips head screwdriver.
- Eject the PXI instrument from the chassis slot using the injector/ejector handle.
- Slide the PXI instrument out of the chassis.
- Slide the new PXI instrument in the open chassis slot.
- Lock the PXI instrument in the chassis slot using the injector/ejector handle.
- Screw in the captive screws on the PXI instrument using the #1 Phillips head screwdriver.
- Connect the appropriate cables to the PXI instrument.
Related Information
- Submitting a Request for RMA, Repair Services, or Replacement Parts
- Maintenance
Perform the following maintenance procedures to ensure the Data Record System operates correctly.