NI 9219 Calibration Procedure
- Updated2025-02-28
- 5 minute(s) read
NI 9219 Calibration Procedure
Calibrate the NI 9219 using Calibration Executive.
Calibration Executive Procedure
Features
Verify Only ![]() |
Verify & Adjust ![]() |
Manual Mode ![]() |
Optional Adjust ![]() |
Approximate Test Time
Verify Only | Verify & Adjust |
---|---|
50 minutes | 150 minutes |
Test Equipment
The following table lists the test equipment required for calibrating NI 9219 modules.
Instrument | Recommended Model | Requirements |
---|---|---|
Calibrator | Fluke 5520 | Voltage Mode
Current Mode
Resistance Mode
|
DMM (x2) | PXI-4071 | Voltage Accuracy DMM 1:
DMM 2:
Resistance Accuracy
Current Accuracy
|
cDAQ Chassis | cDAQ-9178 | — |
Additional Equipment
The following table lists additional equipment that is required for the calibration fixture.
Equipment | Recommended Model | Requirements |
---|---|---|
Switch Matrix | PXI/PXIe-2737 | Switch required for this fixture. |
MIO 63xx Model Device | PXIe-6341 | 8 Digital I/O lines capable of driving 24 mA of current each. |
CAL-9219 | P/N 785860-01 | NI 9219 Calibration Fixture |
Cable Assembly | SH37F-OM6/4 (P/N 146907-01) | 9219 Calibration Cable Kit |
Switch Cable | SH160DIN-SH160DIN (P/N 782417-02) | 160-pin DIN to 160-pin DIN |
Shielded Cable | SHC68-68-EPM (P/N 192061-01) | 68-position D-type to 68-position VHDCI Offset, 1 m cable |
Banana Cables (x10) | — | — |
Legacy Switch Fixture
Prior to version 4.5, Calibration Executive supported a legacy switch fixture. The new switch fixture is recommended; however, if you use the legacy switch fixture, the following equipment is required.
Equipment | Recommended Model | Requirements |
---|---|---|
Switch Matrix | PXI/PXIe-2529 (x2) and PXI-2503 | Switches required for this fixture. |
Switch Fixture | Custom Switch Fixture | Contact NI and reference NI P/N 539725A-025. |
Banana Cables (x10) | — | — |
Test Conditions
The following setup and environmental conditions are required to ensure the NI 9219 meets published specifications.
- Use the switches and switch fixture provided by NI to calibrate the NI 9219 using Calibration Executive.
- Verify that all connections to the device are secure.
- Maintain an ambient temperature of 23 °C ±5 °C. The device temperature will be greater than the ambient temperature.
- Keep relative humidity below 80%.
- Allow a warm-up time of at least 10 minutes to ensure that the NI 9219 measurement circuitry is at a stable operating temperature.
Device Setup
- Install the NI 9219 in slot 8 of the cDAQ-9178 chassis. Leave slots 1 through 7 in chassis empty.
- Configure the hardware using Measurement & Automation Explorer (MAX).
- Launch the Calibration Executive procedure and complete the setup wizard.
- Connect the calibration fixture to the system. Refer to the appropriate setup instructions below, based on the fixture you are using.
CAL-9219 Connection Setup
Complete the following steps before running the Calibration Executive procedure.
- Connect the NI 2737 switch to the "Switch 2" connector on the fixture shown below using the 160-pin DIN cable.
- Connect the NI 6341 device to the "Switch 1" connector using the SHC68-68-EPM
cable.
- Connect the NI 9219 to the "Device Under Test" connector using the NI 9219 Fixture Cable. Refer to the labels on the cable to determine which channels on the NI 9219 to connect.
- Connect the banana cables to the two DMMs and calibrator according to the
associated labels on the CAL-9219 calibration fixture. Keep track of
DMM 1 and DMM 2; these need to
match the configuration in Calibration Executive.
Legacy Switch Fixture
Although NI recommends that you upgrade to the CAL-9219 fixture and PXIe-2737 switches, Calibration Executive still supports the legacy switch fixture. Complete the following steps to set up the legacy fixture.
- Connect the two TB-2636 terminal blocks to the two NI 2529 switches. Make sure to note which switch is connected to TB1 and TB2; these must correlate with NI 2529 #1 and NI 2529 #2 when configuring the devices during setup.
- Connect the TB-2605 (TB3) to the NI 2503.
- Connect the banana cables to their respective instruments based on their labels. Banana connectors that share the same label should be stacked at the instrument. Be sure to configure the two DMMs correctly during setup.
- Connect the four NI 9219 connectors to their respective channels on the NI 9219 under test.
Switch
Maintenance
To ensure proper operation of the switches used in this procedure, you must periodically test the performance of your switches.
If you use the recommended fixture with the NI 2737 switch matrix, you can use the NI Switch Health Center periodically to confirm that the switch is continuing to operate as expected. The NI Switch Health Center can be found in program files, or you can right-click the NI 2737 switch device in NI-MAX to access the NI Switch Health Center.
If you use the legacy fixture with the NI 2737 switch matrix with the NI 2529 and PXI-2503 switches, refer to the NI 2503/2529 Switch Performance Test for instructions on validating the performance of the switches used in this procedure.
Test Limit Equations
The test limits in this procedure are derived from the published specifications found in the NI 9219 Calibration Procedure.
In This Section
Related Information
- Launching a Calibration Procedure
- NI 2503/2529 Switch Performance Test
Test the performance of the PXI-2503 and PXI/PXIe-2529 using Calibration Executive.